SYSTEM OVERVIEW |
System overview:
|
Medical Grade tower UPS system includes built
in isolation transformer with faraday shield, line interactive voltage
regulation. AC surge suppression, RS-232 & USB monitoring ports
with complete software and cabling, plus hospital-grade plug and
outlets. Runtime is expandable via separate purchase BP36V27 external
battery packs. UL 60601-1 listed as medical electrical equipment with
less than 300 microamps combined leakage current for UPS and connected
loads.
|
Voltage compatibility:
|
120V AC
|
Frequency compatibility:
|
60 Hz
|
OUTPUT |
Output volt amp capacity (VA):
|
1000
|
Output watt capacity (watts):
|
750
|
Output nominal voltage:
|
120V AC
|
Output voltage regulation:
|
LINE MODE: Sine wave line voltage 120V (-18% /
+8%)
BATTERY MODE: PWM sine wave output 115V +/-5%
|
Output frequency regulation:
|
LINE MODE: Passes line frequency of 60Hz +/-10%
BATTERY MODE: Inverter output regulated to 60Hz +/-0.5Hz
|
Outlet quantity / type:
|
4 NEMA 5-15R HG
|
Overload protection:
|
Resettable circuit breaker
|
INPUT |
Maximum input amps:
|
9.2A / 860 watts
|
Input connection type:
|
NEMA 5-15P HG
|
Input cord length:
|
6 ft., 14 gauge
|
Recommended electrical service:
|
15A 120V
|
BATTERY |
Full load runtime:
|
16 min. (1000VA)
|
Half load runtime:
|
34 min. (500VA)
|
Expandable battery runtime:
|
Supports extended runtime with optional BP36V27
external battery pack (multi-pack compatible)
|
DC system voltage:
|
36V DC
|
Typical battery lifespan:
|
3-6 years, depending on usage
|
Battery recharge rate:
|
2-4 hours to 90%
|
VOLTAGE REGULATION |
Voltage regulation description:
|
Line interactive UPS offers 2 boost levels of
brownout correction and a single level of overvoltage adjustment to
maintain usable, computer grade output without consuming battery power
over an input voltage range of 81 to 143V AC.
|
Overvoltage correction:
|
Input voltages between 126 and 143V AC are
reduced by 9%
|
Direct pass through:
|
Input voltages between 105 and 125V AC are
passed through unchanged
|
Brownout correction:
|
Input voltages between 96 and 106V AC are
boosted by 10%
|
Severe brownout correction:
|
Input voltages between 81 and 95V AC are
boosted by 21%
|
LEDS ALARMS & SWITCHES |
Front panel LEDs:
|
Yes, includes 5 front-panel indicator LEDs to
show battery charge level, AC line voltage levels, automatic voltage
regulation status, UPS load level and replace battery warning
|
Alarms:
|
3 function audible alarm indicates power
failure conditions (4 short beeps every 10 seconds), overload
(continuous short beeps) and low battery (continuous tone)
|
Switches:
|
Yes, includes 2 front panel mounted push-button
switches for system enable, self-test and alarm cancel functions
|
SURGE / NOISE SUPPRESSION |
AC surge suppression:
|
480 joules
|
AC suppression response time:
|
Instantaneous
|
EMI / RFI AC noise suppression:
|
Yes
|
PHYSICAL |
Shipping weight (lbs):
|
49.2
|
Shipping weight (kg):
|
22.4
|
Unit weight (lbs):
|
45.2
|
Unit weight (kg):
|
20.5
|
Unit Dimensions (HWD/in):
|
12.5 x 7.5 x 9.5
|
Unit Dimensions (HWD/cm):
|
31.75 x 19.1 x 24.2
|
Shipping Dimensions (HWD/in):
|
18.25 x 11.75 x 13.5
|
Shipping Dimensions (HWD/cm):
|
46.4 x 29.9 x 34.3
|
Material of construction:
|
PVC
|
Form factors supported:
|
Tower
|
Cooling method:
|
UPS cooling via rear-exit fan
|
Battery Access:
|
Battery access door allows hot swap battery
replacement without powering connected equipment off
|
ENVIRONMENTAL |
Operating Temperature:
|
+32 to +104 degrees Fahrenheit
0 to +40 degrees Celsius
|
Storage Temperature:
|
+5 to +122 degrees Fahrenheit
-15 to +50 degrees Celsius
|
Relative Humidity:
|
0 to 95%, non-condensing
|
COMMUNICATIONS |
Network monitoring port:
|
Includes 2 built-in monitoring ports, 1 USB
& 1 DB9
|
Software and cabling included:
|
PowerAlert UPS monitoring software is included
on CD-ROM for all common network and standalone operating systems.
Ships with 2 cables, 1-USB to USB & 1-DB9 to DB9
|
LINE / BATTERY TRANSFER |
Transfer time from line power to
battery mode:
|
2-4 milliseconds
|
Low voltage transfer to battery power:
|
81V (resets to line power as voltage increases
to 85V)
|
High voltage transfer to battery power:
|
143V (resets to line power as voltage decreases
to 139V)
|
CERTIFICATIONS |
Certifications:
|
UL1778 (USA, Series AGSM1200PSR3HG), UL 60601-1
(Medical), cUL (Canada), Class A (Emissions)
|
WARRANTY |
Optional coverage:
|
3-year, 5-year, next day and on-site warranty
coverage available in many areas; contact Tripp Lite
|
Product warranty:
|
2 year product warranty
|
SPECIAL FEATURES |
Cold Start:
|
Yes, inverter can be cold started to enable
temporary AC output during a power failure
|
Appearance:
|
Attractive gray tower UPS
|